The digital turbidity electrode adopts the 90° scattered light method, near infrared light source, eliminates chromaticity interference, conforms to ISO7027 standard, the LED emits near infrared beam to the sample at a certain angle, the beam will be scattered when it encounters the suspended matter in the sample in the process of transmission, and the detector, which is set at 90° from the incident light, receives the scattered light, the concentration of suspended matter in the sample is directly proportional to the intensity of the scattered light, and the concentration of suspended matter in the sample and the turbidity of the sample can be calculated by determining the intensity of the scattered light. The concentration of suspended solids in the sample is directly proportional to the intensity of the scattered light, so that the concentration of suspended solids in the sample and the turbidity can be calculated by measuring the intensity of the scattered light.
The digital turbidity electrode adopts the 90° scattered light method, near infrared light source, eliminates chromaticity interference, conforms to ISO7027 standard, the LED emits near infrared beam to the sample at a certain angle, the beam will be scattered when it encounters the suspended matter in the sample in the process of transmission, and the detector, which is set at 90° from the incident light, receives the scattered light, the concentration of suspended matter in the sample is directly proportional to the intensity of the scattered light, and the concentration of suspended matter in the sample and the turbidity of the sample can be calculated by determining the intensity of the scattered light. The concentration of suspended solids in the sample is directly proportional to the intensity of the scattered light, so that the concentration of suspended solids in the sample and the turbidity can be calculated by measuring the intensity of the scattered light.
Product Features
- Low Turbidity Monitoring
Can be used for low turbidity monitoring (<1NTU, with flow cell).
- Automatic cleaning
Automatic cleaning function with wiper.
- Built-in temperature sensor
Built-in temperature sensor
- Multiple output methods
Supports MODBUS/RS-485, 4-20mA current interface output.
- Rugged and waterproof
Corrosion-resistant shell, IP68, can work underwater for a long time.
- Wide range
Wide testing range (0~3000NTU).
Product Features
- Low Turbidity Monitoring
Can be used for low turbidity monitoring (<1NTU, with flow cell).
- Automatic cleaning
Automatic cleaning function with wiper.
- Built-in temperature sensor
Built-in temperature sensor
- Multiple output methods
Supports MODBUS/RS-485, 4-20mA current interface output.
- Rugged and waterproof
Corrosion-resistant shell, IP68, can work underwater for a long time.
- Wide range
Wide testing range (0~3000NTU).
Name | PT-DZ09-20 | PT-DZ09-200 | PT-DZ09-1000 | PT-DZ09-3000 |
Range | 0~20NTU | 0~200NTU | 0~1000NTU | 0~3000NTU |
Resolution (NTU) | 0.01、1 | 0.1、1 | ||
Accuracy | ≤5NTU,≤±0.3NTU; >5NTU,≤±6% | ≤10NTU,≤±0.5NTU; >10NTU,≤±8% | ||
Operating Temperature | 5~40℃ | |||
Maximum Depth | Underwater 5 metres | |||
Digital Interface | MODBUS /RS-485 | |||
Analogue Interface | 4-20mA | |||
Power supply | 12VDC±20%, 1A | |||
Rated power | Normal power: <0.5W Start cleaning brush: <3W | |||
Calibration method | One or two point calibration | |||
Protection class | IP68 | |||
Mounting method | 3/4″NPT thread Flow-through pool mounting, submerged mounting | |||
Dimensions | φ54×150mm | |||
Flow Cell (Optional) | 350×117×198mm | |||
Cable Length | 5 metres (default) | |||
Electrode Material | 316 stainless steel |
Name | PT-DZ09-20 | PT-DZ09-200 | PT-DZ09-1000 | PT-DZ09-3000 |
Range | 0~20NTU | 0~200NTU | 0~1000NTU | 0~3000NTU |
Resolution (NTU) | 0.01、1 | 0.1、1 | ||
Accuracy | ≤5NTU,≤±0.3NTU; >5NTU,≤±6% | ≤10NTU,≤±0.5NTU; >10NTU,≤±8% | ||
Operating Temperature | 5~40℃ | |||
Maximum Depth | Underwater 5 metres | |||
Digital Interface | MODBUS /RS-485 | |||
Analogue Interface | 4-20mA | |||
Power supply | 12VDC±20%, 1A | |||
Rated power | Normal power: <0.5W Start cleaning brush: <3W | |||
Calibration method | One or two point calibration | |||
Protection class | IP68 | |||
Mounting method | 3/4″NPT thread Flow-through pool mounting, submerged mounting | |||
Dimensions | φ54×150mm | |||
Flow Cell (Optional) | 350×117×198mm | |||
Cable Length | 5 metres (default) | |||
Electrode Material | 316 stainless steel |