TO MAKE THE WORLD CLOSER TO THE TRUTH
Availability: | |
---|---|
Quantity: | |
Pure-R
TOP
Category | Pure-R (Basic) | Pure-RUF (Eliminating Endotoxin) | Pure-RUV (Low TOC) | Pure-RFV (Synthesizing) |
Output-RO Water | 20L/Hour | |||
Output - Deionized Water | 2.0L/Min (When tank is full) | |||
Feed Water Requirements | Tap water conductivity <400µs, 5-45°C, 1.0-4.0Kg/cm³ (When TDS>400µs, optional pretreatment is suggested) | |||
Ultrapure Water Quality | ||||
Resistivity (25°C) | 18.28 MΩ·cm | |||
Heavy Metal Ion | <0.01 ppb | |||
TOC | <10 ppb | <10 ppb | <3 ppb | <3 ppb |
Bacteria | <0.01 cfu/ml | |||
Particle (>0.22µm) | <1/ml | |||
Pyrogen (Endotoxin) | N/A | <0.001 EU/ml | N/A | <0.001 EU/ml |
RNases | N/A | <1 pg/ml | N/A | <1 pg/ml |
DNases | N/A | <5 pg/ml | N/A | <5 pg/ml |
RO Water Quality | ||||
Conductivity | 1#RO water,conductivity=source water conducivity ×5%;2*staged RO water:1-5μs/cm(conductivity of source water×2%) | |||
Organic Rejection Rate | >99% (MW >200 Dalton) | |||
Particles/Bacteria Rejection Rate | >99% | |||
Dimensions (W×D×H) | 540×390×640 mm | |||
Weight | ~30 kg | |||
Voltage/Frequency | 100-240V, 50/60Hz | |||
Power | 72W/120W | |||
Standard Configuration | Main system (including 1 full set of cartridges) + 2 built-in tanks (PE and Pressure tank) + accessories bag |
Category | Pure-R (Basic) | Pure-RUF (Eliminating Endotoxin) | Pure-RUV (Low TOC) | Pure-RFV (Synthesizing) |
Output-RO Water | 20L/Hour | |||
Output - Deionized Water | 2.0L/Min (When tank is full) | |||
Feed Water Requirements | Tap water conductivity <400µs, 5-45°C, 1.0-4.0Kg/cm³ (When TDS>400µs, optional pretreatment is suggested) | |||
Ultrapure Water Quality | ||||
Resistivity (25°C) | 18.28 MΩ·cm | |||
Heavy Metal Ion | <0.01 ppb | |||
TOC | <10 ppb | <10 ppb | <3 ppb | <3 ppb |
Bacteria | <0.01 cfu/ml | |||
Particle (>0.22µm) | <1/ml | |||
Pyrogen (Endotoxin) | N/A | <0.001 EU/ml | N/A | <0.001 EU/ml |
RNases | N/A | <1 pg/ml | N/A | <1 pg/ml |
DNases | N/A | <5 pg/ml | N/A | <5 pg/ml |
RO Water Quality | ||||
Conductivity | 1#RO water,conductivity=source water conducivity ×5%;2*staged RO water:1-5μs/cm(conductivity of source water×2%) | |||
Organic Rejection Rate | >99% (MW >200 Dalton) | |||
Particles/Bacteria Rejection Rate | >99% | |||
Dimensions (W×D×H) | 540×390×640 mm | |||
Weight | ~30 kg | |||
Voltage/Frequency | 100-240V, 50/60Hz | |||
Power | 72W/120W | |||
Standard Configuration | Main system (including 1 full set of cartridges) + 2 built-in tanks (PE and Pressure tank) + accessories bag |