PICP-6800 series
TOP
The PICP-6800 inductively coupled plasma emission spectrometer is developed by our company after years of technical accumulation. It is used to determine the content of trace and trace metal elements or non-metal elements in various substances (soluble in hydrochloric acid, nitric acid, hydrofluoric acid, etc.), with high automation, easy operation, stability and reliability. At present, instruments are widely used in various fields such as rare earth, geology, metallurgy, chemical engineering, environmental protection, clinical medicine, petroleum products, semiconductors, food, biological samples, criminal science, agricultural research, etc.
A continuously adjustable solid-state RF power supply with one key ignition and fully automatic impedance matching
Excellent optical system, constant temperature in the light chamber, intelligent blowing
Multi channel peristaltic pump injection ensures uniform instrument injection and stable waste liquid discharge system
High degree of automation, real-time monitoring of status, and abnormal chain protection
The software has powerful functions, complete database management, user-friendly interface design, and multitasking operation
Wide measurement range, analysis from trace to constant, dynamic linear range of 5-6 orders of magnitude.
The PICP-6800 inductively coupled plasma emission spectrometer is developed by our company after years of technical accumulation. It is used to determine the content of trace and trace metal elements or non-metal elements in various substances (soluble in hydrochloric acid, nitric acid, hydrofluoric acid, etc.), with high automation, easy operation, stability and reliability. At present, instruments are widely used in various fields such as rare earth, geology, metallurgy, chemical engineering, environmental protection, clinical medicine, petroleum products, semiconductors, food, biological samples, criminal science, agricultural research, etc.
A continuously adjustable solid-state RF power supply with one key ignition and fully automatic impedance matching
Excellent optical system, constant temperature in the light chamber, intelligent blowing
Multi channel peristaltic pump injection ensures uniform instrument injection and stable waste liquid discharge system
High degree of automation, real-time monitoring of status, and abnormal chain protection
The software has powerful functions, complete database management, user-friendly interface design, and multitasking operation
Wide measurement range, analysis from trace to constant, dynamic linear range of 5-6 orders of magnitude.
Technical indicators of solid-state power supply | |
Circuit type | Inductive feedback self-excited oscillation circuit, coaxial cable output, matching tuning, power feedback closed-loop automatic control |
Frequency Stability | <0.1% |
Output power | 800W-1200W |
Output power stability | <0.3% |
Specification of argon flowmeter and carrier gas pressure gauge | |
Plasma gas flowmeter | (100-1000)L/h(1.6-16L/min) |
Carrier gas stabilizing valve | (0-0.4MPa) |
Cooling circulating water | Water temperature 20-25 ℃, flow rate>5L/min, water pressure>0.1MPa |
Technical specifications of monochromator | |
Light path | Czerny-Turner |
Grating specifications | Ion etched holographic grating with a line density of 3600 lines/mm (optional line density of 2400 lines/mm) |
Resolution | ≤ 0.007nm (3600 lines) ≤ 0.015nm (2400 lines) |
Scanning wavelength range | 3600 lines/mm scanning wavelength range: 190-500nm 2400 lines/mm scanning wavelength range: 190-800nm |
Exit slit | 12um incident slit: 10um |
Technical specifications of photoelectric converters | |
Specification of photomultiplier tube | R293 or R298 |
Technical specifications of the whole machine | |
Scanning wavelength range | 195nm~500nm (3600L/mm grating) 195nm~800nm (2400L/mm grating) |
Repeatability (i.e. short-term stability) | Relative standard deviation RSD ≤ 1.5% |
Stability | Relative standard deviation RSD ≤ 2% |
Technical indicators of solid-state power supply | |
Circuit type | Inductive feedback self-excited oscillation circuit, coaxial cable output, matching tuning, power feedback closed-loop automatic control |
Frequency Stability | <0.1% |
Output power | 800W-1200W |
Output power stability | <0.3% |
Specification of argon flowmeter and carrier gas pressure gauge | |
Plasma gas flowmeter | (100-1000)L/h(1.6-16L/min) |
Carrier gas stabilizing valve | (0-0.4MPa) |
Cooling circulating water | Water temperature 20-25 ℃, flow rate>5L/min, water pressure>0.1MPa |
Technical specifications of monochromator | |
Light path | Czerny-Turner |
Grating specifications | Ion etched holographic grating with a line density of 3600 lines/mm (optional line density of 2400 lines/mm) |
Resolution | ≤ 0.007nm (3600 lines) ≤ 0.015nm (2400 lines) |
Scanning wavelength range | 3600 lines/mm scanning wavelength range: 190-500nm 2400 lines/mm scanning wavelength range: 190-800nm |
Exit slit | 12um incident slit: 10um |
Technical specifications of photoelectric converters | |
Specification of photomultiplier tube | R293 or R298 |
Technical specifications of the whole machine | |
Scanning wavelength range | 195nm~500nm (3600L/mm grating) 195nm~800nm (2400L/mm grating) |
Repeatability (i.e. short-term stability) | Relative standard deviation RSD ≤ 1.5% |
Stability | Relative standard deviation RSD ≤ 2% |